MEAN WELL GEM40I05-P1J delivers 5 V at a rated 5 A, with a catalogue-rated maximum output of 25 W. This Class II medical adapter has an interchangeable AC plug interface and a centre-positive P1J DC lead; the AC plug kit is sold separately.
Product Datasheet
Download MEAN WELL GEM40I Datasheet (PDF)
The GEM40I section is on PDF pages 17, 18, 19. Refer to the complete mechanical drawings and notes when planning installation.
Model and configuration
| Full ordering code |
GEM40I05-P1J |
| Series and product |
GEM40I single-output medical AC/DC adapter with interchangeable AC plugs. |
| Code meaning |
GEM: medical adapter series; 40: nominal family class; I: interchangeable AC plug; 05: 5 V output; P1J: standard centre-positive tuning-fork DC plug. |
| Safety model |
GEM40I05 |
| Supply configuration |
P1J DC output lead is the standard configuration. The interchangeable AC plug kit is sold separately. |
Output specifications
| DC output voltage |
5 V |
| Rated current / current range |
5 A rated; 0 to 5 A range. |
| Catalogue-rated maximum power |
25 W. Select the load within both the rated current and rated power limits. |
| Ripple and noise, maximum |
100 mVp-p |
| Voltage tolerance |
±5.0% |
| Line / load regulation |
±1.0% / ±5.0% |
| Typical efficiency |
84% |
| No-load consumption / efficiency level |
<0.1 W; Level VI. EISA 2007 / DoE and EU ErP. |
Input and start-up
| Input range / frequency |
80–264 VAC or 113–370 VDC; 47–63 Hz. |
| AC input current |
1 A at 115 VAC; 0.6 A at 230 VAC. |
| Maximum inrush current |
30 A at 115 VAC; 65 A at 230 VAC. |
| Set-up / rise / hold-up time |
500 / 30 / 16 ms at full load, at both 115 VAC and 230 VAC. |
| Input class |
Class II, without an earth pin. |
| Touch current |
<100 µA at 264 VAC. |
Protection
| Short circuit and overload |
Short-circuit protection is listed. Overload: 105–160% of rated output power; hiccup mode recovers automatically after the fault is removed. |
| Overvoltage |
110–140% of rated output voltage; zener-diode clamping with the output-short protection mode stated in the datasheet. |
Temperature and input derating
| Working ambient range |
-25 to +70°C, subject to the load curve. |
| Thermal curve |
100% load from -25 to +40°C; reduce along the plotted line to 50% at +70°C. |
| Low-input-voltage curve |
80% load at 80 VAC, increasing to 100% at 115 VAC; 100% from 115 to 264 VAC. |
| Working humidity |
20–90% RH, non-condensing. |
| Storage |
-25 to +85°C; 10–95% RH. |
| Temperature coefficient |
±0.03%/°C from 0 to 40°C. |
| Vibration |
10–500 Hz, 2 G; 10 minutes per cycle, 60 minutes each along X, Y and Z. |
Apply both temperature and input-voltage derating when selecting the permitted load. The wider working range does not mean full-load operation at every temperature.
Medical safety and system integration
| Isolation / patient environment |
2 × MOPP between primary and secondary. Suitable for BF applications with appropriate system consideration. |
| Safety approvals in this catalogue |
ANSI/AAMI ES60601-1 / 60601-1-11 and TUV EN60601-1 / 60601-1-11 approved. |
| Input-to-output withstand |
5656 VDC. |
| Insulation resistance |
Input to output: 100 MΩ at 500 VDC, 25°C and 70% RH. |
| EMC emissions |
EN55011 Class B; EN61000-3-2,3; FCC PART15 / CISPR22 Class B. |
| EMC immunity |
EN61000-4-2,3,4,5,6,8,11; light-industry level, criteria A. |
The adapter's medical isolation and BF suitability must be assessed within the complete equipment. The finished system must be checked again for EMC compliance.
Typical applications listed for this family include blood-glucose and blood-pressure meters, nebulizers, inhalers, portable medical devices and sleep-apnea devices.
Dimensions, DC connection and installation
| Body dimensions |
75.5 × 39.1 × 56.2 mm (L × W × H). |
| Construction |
Fully enclosed 94V-0 flame-retardant plastic case; wall-mounted adapter format. |
| Standard P1J DC plug |
Tuning-fork type; 2.1 mm ID × 5.5 mm OD; insertion length 11 ±0.5 mm. Centre positive, outside negative. |
| Output cable for this model |
UL1571, 14 AWG, 1000 mm (+150/-0 mm length tolerance). |
| Single-unit catalogue weight |
200 g, the individual weight listed in PACKING; a reference value for this catalogue configuration. |
Fit the required regional AC plug to the interchangeable interface and match the equipment's DC connector and polarity. Follow the full mechanical drawing and the manufacturer's installation instructions.
Compatible options and ordering
| AC plug-AU2 |
Australian type; compatible interchangeable AC plug for GEM18I / GEM30I / GEM40I. |
| AC plug-UK2 |
U.K. type; compatible interchangeable AC plug for GEM18I / GEM30I / GEM40I. |
| AC plug-EU2 |
European type; compatible interchangeable AC plug for GEM18I / GEM30I / GEM40I. |
| AC plug-US2 |
US type; compatible interchangeable AC plug for GEM18I / GEM30I / GEM40I. |
| AC plug-MIX2 |
Mixed set of all four listed types; compatible interchangeable AC plug for GEM18I / GEM30I / GEM40I. |
| Optional locking DC plug |
P2S; SWITCHCRAFT S761K plug equivalent. Specify this option when ordering; the present -P1J model uses P1J. |
| Other plug or cable configurations |
Available by customer request; confirm the exact configuration with the supplier. |
The AC plug kit is sold separately. P2S is an alternative DC connector configuration, not an additional plug included with the standard P1J unit.
Reliability and measurement conditions
| Burn-in / manufacturer warranty |
100% full-load burn-in test; the manufacturer's datasheet lists a 3-year warranty. |
| Life rating |
3 years at 100% load, 40°C, 12 operating hours per day. |
| MTBF |
400,000 hours minimum, MIL-HDBK-217F at 25°C. |
| Default measurement conditions |
230 VAC input, rated load, 25°C and 70% RH, unless otherwise stated. |
| Voltage set point |
Measured at the plug terminals at 50% load. |
| Ripple measurement |
20 MHz bandwidth; 12-inch twisted pair terminated with 0.1 µF and 47 µF capacitors. |
| Regulation definitions |
Tolerance includes set-up tolerance, line regulation and load regulation. Line regulation is measured from low to high line at rated load; load regulation from 10% to 100% rated load. |
| Catalogue revision |
2016-07-11 |